FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 3091073 · Received April 30, 2013

Report

Report Number
2221819-2013-00872
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
March 27, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS SUA, INC.
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

SERVICE REPRESENTATIVE EVALUATED THE SYSTEM AND FOUND EXCESSIVE HEAT IN THE RACK ROOM. CUSTOMER WAS PROVIDED WITH A LOANER SYSTEM, WHICH INCLUDED THE SERVER, WIRELESS TRANSCEIVER, TIM, AND SWITCH, WHICH IT IS BEING RETURNED TO MINDRAY'S NATIONAL REPAIR CENTER FOR REPAIR.

Description of Event or Problem · 1

CUSTOMER REPORTED THE PANORAMA CENTRAL STATION SHUT DOWN, WHICH MAY HAVE RESULTED IN LOSS OF AMBULATORY TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186976 PANORAMA CENTRAL STATION CENTRAL MONITORING STATION MHX MINDRAY DS SUA, INC.

Patients

Seq Age Sex Outcome Treatment
1