FDA Adverse Event
Malfunction
Summary report: N
A5 ANESTHESIA SYSTEM
MDR report key: 3091052
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00893
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- April 1, 2013
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- BSZ
- PMA / PMN Number
- 123211
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REPRESENTATIVES FOUND THE O2 HOUSING WAS NOT TIGHTENED COMPLETELY AS THE CAUSE OF THE ISSUE.
Description of Event or Problem · 1
CUSTOMER REPORTED THE A5 ANESTHESIA SYSTEM HAS A SYSTEM LEAK. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 188207 | A5 ANESTHESIA SYSTEM | ANESTHESIA SYSTEM DELIVERY | BSZ | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |