FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 3091043 · Received April 30, 2013

Report

Report Number
2221819-2013-01015
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
March 27, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE ANTENNA. SYSTEM WAS TESTED, CALIBRATED, AND SAFETY CHECKED TO FACTORY SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED AN ISSUE WITH THE ANTENNA ON THE PANORAMA CENTRAL STATION'S WIRELESS TRANSCEIVER (TIM), WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188695 PANORAMA CENTRAL STATION CENTRAL STATION MONITOR MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1