FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3091026 · Received May 3, 2013

Report

Report Number
9611451-2013-00354
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 2, 2013
Report Date
April 4, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICES RETURNED FOR INVESTIGATION: DEVICE 1: MR290V LOT#: 121001; DEVICE 2: MR290V LOT#: 121002; DEVICE 3: MR290V LOT#: 120912. METHOD: THREE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WERE VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: VISUAL INSPECTION REVEALED A BREAK IN THE WATER FEEDSET TUBE AT THE CONNECTION TO THE CHAMBER DOME ON DEVICES 1 AND 3. DEVICE 2 EXHIBITED A BREAK IN THE WATER FEEDSET TUBE WHERE IT IS INSERTED TO THE BAG SPIKE. THE SURFACE OF THE BREAKS ON ALL THREE WATER FEEDSET TUBES WERE ROUGH (NOT SMOOTHLY CUT). A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 121001 AND THREE OTHER COMPLAINTS FOR LOT NUMBER 120912. NO COMPLAINTS OF THIS NATURE WERE IDENTIFIED FOR LOT 121002. CONCLUSION: THE BREAK ON ALL THREE FEEDSET TUBES WAS ROUGH, SUGGESTING THAT THE DAMAGE MAY HAVE OCCURRED AS A RESULT OF THE TUBES BEING PULLED AWAY FROM THE CHAMBERS POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR UNDER TENSION. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE DAMAGE OCCURRED POST PRODUCTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO OUR DISTRIBUTOR THAT THE WATER FEEDSET TUBES ON THREE MR290 AUTOFEED HUMIDIFICATION CHAMBERS WERE DAMAGED. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193885 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 120912

Patients

Seq Age Sex Outcome Treatment
1