ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2013-00421
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 2, 2013
- Report Date
- April 3, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- MMI
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) REPLACED THE TRANSDUCER AND PERFORMED ALIGNMENTS, AND THE ULTRASONICS VOLTAGES WERE SET TO INSTRUMENT SPECIFICATION. TO VERIFY PROPER SYSTEM OPERATION, THE FSE PERFORMED MULTIPLE PATIENTS PROGRAMING, TWO REPETITIONS OF TROPONIN I. NO FALSE HIGH RESULTS WERE NOTED. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. ALL RELATED MEDWATCH REPORTS ASSOCIATED WITH THIS INCIDENT: 2122870-2013-00421, 2122870-2013-00422, 2122870-2013-00423.
THE CUSTOMER REPORTED ERRONEOUSLY ELEVATED TROPONIN I (ACCESS ACCUTNI) RESULTS, WITHIN THE RISK STRATIFICATION AND ABOVE THE ACUTE MYOCARDIAL INFARCTION (AMI) LIMIT, FOR SEVEN PATIENTS, ON THREE SEPARATE DAYS INVOLVING THE ACCESS 2 IMMUNOASSAY SYSTEM USED IN CONJUNCTION WITH THE ACCESS ACCUTNI ASSAY AND CALIBRATOR. THE CUSTOMER STATED EACH PATIENT'S SAMPLE PRODUCED DIFFERENT RESULTS AFTER ANALYZING THE SAMPLE SEVERAL TIMES ON THE ACCESS 2 SYSTEM. SUBSEQUENT ANALYSES OF THE PATIENTS' SAMPLES, ON AN ALTERNATE METHODOLOGY, PRODUCED NEGATIVE RESULTS. THE CUSTOMER STATED THE SAMPLES WERE REPEATED DUE TO DELTA CHECK WITH PREVIOUS PATIENT RESULTS. THE ERRONEOUS PATIENT RESULTS WERE NOT RELEASED OUT OF THE LABORATORY. THERE WAS NO PATIENT CONSEQUENCE ASSOCIATED WITH THIS EVENT. THE CUSTOMER CONTINUED TO ANALYZE PATIENT SAMPLES ON THE ALTERNATE METHODOLOGY. QUALITY CONTROL (QC) IS PERFORMED ONCE EVERY 24 HOURS. QUALITY CONTROL WAS WITHIN SPECIFICATION PRIOR TO THE EVENT; QC WAS NOT PERFORMED AFTER THE EVENT. THE PATIENTS' SAMPLES WERE COLLECTED BY THE EMERGENCY ROOM (ER) STAFF, IN THE ER. THE SAMPLES WERE COLLECTED IN 4.0 ML LITHIUM HEPARIN NON-GEL PLASMA TUBES AND CENTRIFUGED AT 3,600 RPM (ROTATIONS PER MINUTE) FOR SEVEN MINUTES FROM THE PRIMARY TUBES, AT ROOM TEMPERATURE. SAMPLES ARE STORED IN REFRIGERATION. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THIS IS REPORT ONE OF THREE REFERENCING THE EVENT DATE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 192499 | ACCESS® 2 IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | MMI | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |