Report

Report Number
2432235-2013-00161
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 7, 2013
Report Date
April 8, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS
Product Code
MMI
PMA / PMN Number
K041133
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. AFTER EVALUATION OF THE INSTRUMENT AND INSTRUMENT DATA, THE FSE DID NOT FIND AN INSTRUMENT MALFUNCTION. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED TROPONIN RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.

Description of Event or Problem · 1

A DISCORDANT, FALSELY ELEVATED TROPONIN RESULT WAS OBTAINED ON ONE PATIENT SAMPLE ON AN ADVIA CENTAUR XP INSTRUMENT. THE SAMPLE HAD BEEN RUN IN DUPLICATE, AND THE RESULTS DID NOT MATCH. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS RERUN IN DUPLICATE ON AN ALTERNATE ADVIA CENTAUR XP INSTRUMENT, AND THE RESULTS MATCHED THE LOWER RESULT FROM THE INITIAL RUN. THE RERUN RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED TROPONIN RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191433 ADVIA CENTAUR XP IMMUNOASSAY ANALYZER MMI SIEMENS HEALTHCARE DIAGNOSTICS ADVIA CENTAUR XP