ADVIA CENTAUR XP
Report
- Report Number
- 2432235-2013-00161
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 7, 2013
- Report Date
- April 8, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS
- Product Code
- MMI
- PMA / PMN Number
- K041133
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. AFTER EVALUATION OF THE INSTRUMENT AND INSTRUMENT DATA, THE FSE DID NOT FIND AN INSTRUMENT MALFUNCTION. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED TROPONIN RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.
A DISCORDANT, FALSELY ELEVATED TROPONIN RESULT WAS OBTAINED ON ONE PATIENT SAMPLE ON AN ADVIA CENTAUR XP INSTRUMENT. THE SAMPLE HAD BEEN RUN IN DUPLICATE, AND THE RESULTS DID NOT MATCH. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS RERUN IN DUPLICATE ON AN ALTERNATE ADVIA CENTAUR XP INSTRUMENT, AND THE RESULTS MATCHED THE LOWER RESULT FROM THE INITIAL RUN. THE RERUN RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED TROPONIN RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191433 | ADVIA CENTAUR XP | IMMUNOASSAY ANALYZER | MMI | SIEMENS HEALTHCARE DIAGNOSTICS | ADVIA CENTAUR XP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |