FDA Adverse Event Malfunction Summary report: N

VISIAN ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 3090831 · Received May 2, 2013

Report

Report Number
2023826-2013-00334
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 9, 2013
Report Date
April 12, 2013
Manufacturer
STAAR SURGICAL COMPANY
Product Code
MTA
PMA / PMN Number
P030016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MANUFACTURER? NO - LENS NOT RETURNED. METHOD: WORK ORDER SEARCH. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY AND WORK ORDER SEARCH, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4). LENS NOT RETURNED.

Additional Manufacturer Narrative · 1

EVALUATION RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC AND ONE HAPTIC TORN. THE LENS WAS RETURNED IN LIQUID. CONCLUSIONS - (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON WAS INSERTING A 13.2MM MICL 13.2 IMPLANTABLE COLLAMER LENS AND THE FRONT HAPTIC TORE. THE LENS WAS PARTIALLY INSERTED AND WAS REMOVED WITH NO PATIENT INJURY, NO INCISION ENLARGEMENT OR SUTURES. THE BACKUP LENS WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191411 VISIAN ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS MTA STAAR SURGICAL COMPANY MICL 13.2 NA

Patients

Seq Age Sex Outcome Treatment
1 INJECTOR MODEL MSI-PF - LOT NUMBER UNK| CARTRIDGE MODEL SFC-45 FP - LOT NUMBER UNK| FOAM TIP PLUNGER MODEL FTP - LOT NUMBER UNK