FDA Adverse Event Malfunction Summary report: N

COULTER® LH 500 HEMATOLOGY ANALYZER

MDR report key: 3090830 · Received May 2, 2013

Report

Report Number
1061932-2013-00796
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 4, 2013
Report Date
April 4, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) DISCOVERED THE SILICON TUBING HAD A HAIR-LINE TEAR WHICH CAUSED THE FLUID TO SPRAY INSIDE THE REAR DILUTER, BEHIND THE BATH ASSEMBLIES, AT THE 4-INCH RESTRICTOR LINE. THE FSE REPLACED THE TORN TUBING AND SEVERAL PINCH VALVE TUBING, IN THE REAR DILUTER PANEL, DUE TO EXCESSIVE WEAR. THE FSE NOTED QUALITY CONTROL (QC) RECOVERED WITHIN THE ACCEPTABLE LIMITS AND BACKGROUND COUNTS AND LATRON CONTROL RESULTS WERE ALSO WITHIN RANGE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY THREE TO FIVE MILLILITERS OF CLEAR FLUID LEAKED INSIDE THE INSTRUMENT DURING LATRON CONTROL ANALYSIS INVOLVING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE OPERATOR WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. NO ERRONEOUS PATIENT RESULTS WERE GENERATED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. ALL QUALITY CONTROL (QC) RESULTS RECOVERED WITHIN THE ACCEPTABLE LIMITS. THE LABORATORY HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191612 COULTER® LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1