OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-09919
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Report Date
- April 8, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Removal / Correction Number
- 2013-IE-017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THE PATIENT'S PRODUCTS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION; HOWEVER, THE EVALUATION PROCESS HAS NOT YET BEEN COMPLETED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 (08/09/2013)-PRODUCT EVALUATION: THE SUBJECT METER WAS RETURNED ON 05/01/2013 AND ANALYSIS COMPLETED ON 08/05/2013 BY LIFESCAN PRODUCT ANALYSIS. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. THE DEVICE MET ALL SPECIFICATIONS FOR TESTING AND NO ISSUES WERE OBSERVED DURING INVESTIGATION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2013, THE LAY USER/PATIENT IN (B)(6) CONTACTED LIFESCAN TO REPORT THE ONE TOUCH VERIO IQ METER WAS GIVING INACCURATELY HIGH READINGS COMPARED TO ANOTHER METER. THE PATIENT OBTAINED THE BLOOD GLUCOSE READING OF 6.8 MMOL/L ON THE REPORTED METER AND A READING OF 1.2 MMOL/L ON THE OTHER METER. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED, THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191900 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR |