FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3090742 · Received May 2, 2013

Report

Report Number
3004209178-2013-93083
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
May 1, 2013
Report Date
May 1, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS MONITORED AND PROGRAMMED WITH SEVERAL BOLUSES AND DELIVERED PROPERLY. BOLUS WAS DELIVERED AND RECORDED PROPERLY IN BOLUS HISTORY. ALL OPERATING CURRENTS TESTED WITHIN SPECIFICATION RANGE. NO BLANK DISPLAY NOTED AND NO BLANK DISPLAY NOTED DURING BOLUS DELIVERY. THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP NOTED DURING OUR VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT IS REPORTED THAT THE INSULIN PUMP STOPPED INSULIN DELIVERY DURING A BOLUS. CHECKED ALARM HISTORY AND DID NOT FIND NO OFF NO POWER ALARMS. CUSTOMER CONCERNED THAT INSULIN PUMP IS STARTING AND STOPPING INSULIN DELIVERY. NOTHING FURTHER REPORTED.