FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3090731 · Received May 2, 2013

Report

Report Number
3004209178-2013-93081
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 8, 2013
Report Date
April 8, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP RECEIVED WITH MISSING END CAP STICKER.

Description of Event or Problem · 1

IT IS REPORTED THAT THE INSULIN ALARMED MOTOR ERROR. EXPLAINED MOTOR ERROR ALARM AND CORRECTED. CUSTOMER STATES THAT THE INSULIN PUMP WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. CUSTOMER STATED THAT SHE CANNOT REWIND THE INSULIN PUMP. ADVISED CUSTOMER TO RETURN THE INSULIN PUMP FOR ANALYSIS. DISCONTINUE USE. NOTHING FURTHER REPORTED.