FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3090731
·
Received May 2, 2013
Report
- Report Number
- 3004209178-2013-93081
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 8, 2013
- Report Date
- April 8, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP RECEIVED WITH MISSING END CAP STICKER.
Description of Event or Problem · 1
IT IS REPORTED THAT THE INSULIN ALARMED MOTOR ERROR. EXPLAINED MOTOR ERROR ALARM AND CORRECTED. CUSTOMER STATES THAT THE INSULIN PUMP WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. CUSTOMER STATED THAT SHE CANNOT REWIND THE INSULIN PUMP. ADVISED CUSTOMER TO RETURN THE INSULIN PUMP FOR ANALYSIS. DISCONTINUE USE. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191801 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |