FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3090730
·
Received May 2, 2013
Report
- Report Number
- 3004209178-2013-93078
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 3, 2013
- Report Date
- April 3, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM THE BASIC OCCLUSION TEST, OCCLUSION AND EXCESSIVE NO DELIVERY TESTS DUE TO PRIME/FILL ANOMALY. THE INSULIN PUMP HAD A SEVERELY SCRATCHED DISPLAY WINDOW, SCRATCHED CASE, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND A BROKEN BELT CLIP SLOT.
Description of Event or Problem · 1
CUSTOMER CALLED AND REPORTED DAMAGE TO THE INSULIN PUMP, IT HAS A CRACK. CUSTOMER ALSO STATED THAT THE DRIVE SUPPORT CAP IS PROTRUDED. ADVISED CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP. ADVISED CUSTOMER THAT INSULIN PUMP WILL NEED TO BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191324 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |