FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3090730 · Received May 2, 2013

Report

Report Number
3004209178-2013-93078
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 3, 2013
Report Date
April 3, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM THE BASIC OCCLUSION TEST, OCCLUSION AND EXCESSIVE NO DELIVERY TESTS DUE TO PRIME/FILL ANOMALY. THE INSULIN PUMP HAD A SEVERELY SCRATCHED DISPLAY WINDOW, SCRATCHED CASE, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND A BROKEN BELT CLIP SLOT.

Description of Event or Problem · 1

CUSTOMER CALLED AND REPORTED DAMAGE TO THE INSULIN PUMP, IT HAS A CRACK. CUSTOMER ALSO STATED THAT THE DRIVE SUPPORT CAP IS PROTRUDED. ADVISED CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP. ADVISED CUSTOMER THAT INSULIN PUMP WILL NEED TO BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191324 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 38 YR