PENUMBRA PX SLIM DELIVERY MICROCATHETER
Report
- Report Number
- 3005168196-2013-00162
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 3, 2013
- Report Date
- April 3, 2013
- Manufacturer
- PENUMBRA, INC.
- Product Code
- DQY
- PMA / PMN Number
- K100826
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
RESULTS: THE FASTENER ON THE DISTAL END OF THE RHV IS NO LONGER ATTACHED TO THE UNIT. HOWEVER, IT CAN BE REATTACHED AND TIGHTENED DOWN TO OPERATE NORMALLY. CONCLUSION: THE COMPLAINT HAS BEEN EVALUATED AND CONFIRMED. THE COMPLAINT STATES THAT THE PHYSICIAN HAD TROUBLE INSERTING THE CATHETER THROUGH THE RHV AND THAT IN ORDER TO ADVANCE THE CATHETER, THE PHYSICIAN ATTEMPTED TO OPEN THE RHV MORE. HOWEVER, WHILE ATTEMPTING TO OPEN THE RHV MORE TO INSERT THE CATHETER THE RHV BROKE. WHEN THE PHYSICIAN ATTEMPTED TO OPEN THE RHV FURTHER, IT WAS TURNED PAST THE POINT IT WAS INTENDED TO AND THE FASTENER CAME FREE OF THE SCREW THREADS. IF A RHV IS FORCED OPEN MORE THAN IT IS DESIGNED TO, IT IS POSSIBLE FOR DAMAGE TO OCCUR TO PRODUCT. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY CONCERNS.
PHYSICIAN OPENED PX SLIM 45 DEGREES TIP CATHETER IN PREPARATION FOR TREATING PATIENT WITH PULMONARY AVM. PHYSICIAN THEN TRIED TO LOAD CATHETER INTO THE ROTATING HEMOSTATIC VALVE (RHV) AND ENCOUNTERED RESISTANCE. HE THEN TWISTED THE RHV TO OPEN IT FURTHER, AND THE RHV BROKE IN HIS HANDS. HE THEN USED A DIFFERENT RHV AND THE SAME CATHETER TO COMPLETE THE CASE. THIS DID NOT CAUSE ANY PATIENT HARM AS THIS WAS RESOLVED BEFORE INSERTING CATHETER INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191200 | PENUMBRA PX SLIM DELIVERY MICROCATHETER | DQY | DQY | PENUMBRA, INC. | F32654 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |