FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3090663 · Received May 2, 2013

Report

Report Number
3008382007-2013-09915
Event Type
Malfunction
Date Received
May 2, 2013
Report Date
April 5, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Removal / Correction Number
2013-IE-017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (6/20/2013) -DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 6/12/2013 WITH THE FOLLOWING FINDINGS: THE METER FAILED TESTING. THERE WAS FOUND TO BE AN OPEN/SHORTED CAPACITOR. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

THE PATIENT'S PRODUCTS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION; HOWEVER, THE EVALUATION PROCESS HAS NOT YET BEEN COMPLETED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF LIFESCAN OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE LAY USER/PATIENT IN THE (B)(4) CONTACTED LIFESCAN TO REPORT THE ONE TOUCH VERIO IQ METER WOULD NOT POWER ON. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING THIS COMPLAINT IS BEING REPORTED.