0.8% SURGISCREEN FOR GEL
Report
- Report Number
- 2250051-2013-00118
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 10, 2013
- Report Date
- May 2, 2013
- Manufacturer
- ORTHO CLINICAL DIAGNOSTICS
- Product Code
- KSZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
OCD PERFORMED RETAINED TESTING, BATCH RECORD REVIEW, AND COMPLAINT BY LOT REVIEW. ALL RESULTS WERE SATISFACTORY. SAMPLE WAS NOT AVAILABLE TO BE SENT TO OCD FOR FURTHER INVESTIGATION. (B)(4).
CUSTOMER CONTACTED CTS TO REPORT FALSE NEGATIVE REACTIONS IN THE MTS GEL SYSTEM WITH A PATIENT SAMPLE CONFIRMED TO HAVE ANTI-E, ANTI-C, AND ANTI-KELL DETECTED IN THE SOLID PHASE METHODOLOGY ON THE ECHO INSTRUMENT. CUSTOMER REPORTED THAT THE PATIENT WAS INITIALLY TESTED ON THE IMMUCOR ECHO AND YIELDED A POSITIVE REACTION. CUSTOMER STATED THAT IT IS THEIR PROTOCOL TO REPEAT ANY POSITIVE ANTIBODY SCREEN BY THE ECHO IN THE MANUAL GEL METHOD. PATIENT SAMPLE WAS THEN TESTED BY THE MANUAL GEL METHOD WITH A 15MIN INCUBATION AND NEGATIVE RESULTS WERE OBSERVED. CTS CONFIRMED ACCEPTABLE QC TO THE 0.8% SURIGSCREEN AND THE ANTI-IGG GEL CARDS. ALL LISTED OCD PRODUCTS HAVE BEEN CONFIRMED TO HAVE NORMAL APPEARANCE AND STORED ACCORDING TO THE RESPECTED IFU. CTS SUGGESTED ANTIGEN TYPING OF THE 0.8% DONORS THAT WERE EXPECTED TO BE POSITIVE FOR THE RESPECTIVE ANTIBODY. CUSTOMER IS UNWILLING TO PERFORM THE SUGGESTED TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191880 | 0.8% SURGISCREEN FOR GEL | REAGENT RED BLOOD CELLS | KSZ | ORTHO CLINICAL DIAGNOSTICS | VSS536 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |