FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3090610 · Received May 2, 2013

Report

Report Number
3004209178-2013-93075
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 30, 2013
Report Date
April 30, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNABLE TO PRIME DURING PRIME ALARM TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. THE INSULIN PUMP PASSED DISPLACEMENT TEST. UNABLE TO PERFORM OCCLUSION, BASIC OCCLUSION AND EXCESSIVE NO DELIVERY ALARMS DUE TO PRIME ANOMALY. MOTOR WAS TESTED. NO MOTOR ANOMALY NOTED. CRACKED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT IS REPORTED THAT THE DRIVE SUPPORT CAP IS PROTRUDED. CUSTOMER'S FATHER ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192946 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722CAB

Patients

Seq Age Sex Outcome Treatment
1