PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-93075
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 30, 2013
- Report Date
- April 30, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
UNABLE TO PRIME DURING PRIME ALARM TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. THE INSULIN PUMP PASSED DISPLACEMENT TEST. UNABLE TO PERFORM OCCLUSION, BASIC OCCLUSION AND EXCESSIVE NO DELIVERY ALARMS DUE TO PRIME ANOMALY. MOTOR WAS TESTED. NO MOTOR ANOMALY NOTED. CRACKED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP NOTED DURING VISUAL INSPECTION.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT IS REPORTED THAT THE DRIVE SUPPORT CAP IS PROTRUDED. CUSTOMER'S FATHER ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 192946 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722CAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |