FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3090531 · Received May 2, 2013

Report

Report Number
2531779-2013-05815
Event Type
Malfunction
Date Received
May 2, 2013
Report Date
April 10, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 07/11/2013: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 06/13/2013 WITH THE FOLLOWING RESULTS: PUMP RETURNED WITH THE AUDIO BOLUS BUTTON COVER DETACHED FROM THE PUMP. UNABLE TO TEST THE AUDIO BOLUS BUTTON DUE TO THE MISSING COVER.THE UP, DOWN, OK AND CONTRAST KEYS ARE RESPONSIVE TO USER PRESSES.. REMOVED THE KEYPAD COVER; CONTAMINATION WAS PRESENT UNDER THE UP, OK, DOWN AND CONTRAST KEY CONTACTS. UNRELATED TO THIS COMPLAINT, A PINKISH CONTRAST WAS OBSERVED ON THE DISPLAY SCREEN. REMOVED THE PUMP COVER AND REPLACED PINK DISPLAY WITH TEST DISPLAY. THE TEST DISPLAY HAS NORMAL CONTRAST WITH NO VISIBLE SIGNS OF DISCOLORATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (BUTTON/KEYPAD ISSUE) ISSUE. THE REPORTER STATED THE AUDIO BOLUS BUTTON IS "LOOSE." THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193059 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 68 YR