LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2013-02110
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 9, 2013
- Report Date
- April 9, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION: WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? CYSTIC DUCT. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? YES. WERE THERE ANY FEEDING ISSUES EXPERIENCED WITH THE DEVICE? YES. WAS THE SURGEON ABLE TO VISUALIZE A CLIP FED INTO THE JAWS PRIOR TO FIRING THE DEVICE? YES. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? NO. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO. WERE ANY UNEXPECTED NOISES HEARD? NO. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? THE DEVICE WAS FIRED OUT OF THE PATIENT AFTER THE INCIDENT. WHAT WERE THE INDICATIONS FOR SURGERY? ASKU. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? YES, ERCP DONE IN (B)(6) 2013 AND STINTED THE COMMON BILE DUCT. THE PATIENT HAD AN ENLARGED CYSTIC DUCT. DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? NO.
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING THE LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE FIRST CLIP FIRED FINE. THE SECOND CLIP ONLY LOADED HALF-WAY INTO THE JAWS. THE DEVICE WAS TAKEN OUT OF THE PATIENT. THE DEVICE WAS FIRED AND THE SECOND CLIP DROPPED OUT OF THE JAWS OF THE DEVICE. THE DEVICE WAS FIRED AGAIN AND THE CLIP DID NOT LOAD. THE DEVICE WAS FIRED AGAIN AND THE THIRD CLIP FINALLY LOADED WITH NO ISSUES. THREE OR FOUR MORE CLIPS WERE USED AFTER THE EVENT AND THEY LOADED AND FIRED APPROPRIATELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193060 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | K9063J |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) |