FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3090259 · Received May 2, 2013

Report

Report Number
2938836-2013-01063
Event Type
Injury
Date Received
May 2, 2013
Date of Event
March 5, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS CAPPED AND REPLACED DUE TO FRACTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191708 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 24 YR Required Intervention