FDA Adverse Event
Malfunction
Summary report: N
QUADRA ASSURA CRT-D, DF4 CONNECTOR
MDR report key: 3090252
·
Received May 2, 2013
Report
- Report Number
- 2938836-2013-01258
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- February 5, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NIK
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED IN THE CLINIC FOR A FOLLOW UP AND POST-PACED T-WAVE OVERSENSING WAS OBSERVED ON A STORED EGM. PROGRAMMING OPTIONS WERE RECOMMENDED. THE DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 192424 | QUADRA ASSURA CRT-D, DF4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | CD3265-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |