FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3090242 · Received May 2, 2013

Report

Report Number
2938836-2013-01062
Event Type
Injury
Date Received
May 2, 2013
Date of Event
March 26, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED IN THE OR FOR LEAD REPLACEMENT DUE TO ELECTRICAL MEASUREMENTS ANOMALY. HIGH PACING LEAD IMPEDANCE, INCREASED CAPTURE THRESHOLD AND EXTERNALIZED CONDUCTORS WERE OBSERVED. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191379 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention