RIATA ACTIVE FIXATION

MDR report key: 3090241 · Received May 2, 2013

Report

Report Number
2938836-2013-01056
Event Type
Injury
Date Received
May 2, 2013
Date of Event
March 6, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

EVALUATION DESCRIPTION: A PARTIAL LEAD WITH THE IS-1 CONNECTOR BOOT PORTION MEASURING 6.1 CM WAS RETURNED FOR ANALYSIS. THE DAMAGE WAS FOUND SUSTAINED DURING SURGICAL PROCEDURE. THE PORTION OF THE LEAD WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD A COMPLETE ANALYSIS COULD NOT BE CONFIRMED. (B)(4).

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT EXTERNALIZED CONDUCTORS WERE OBSERVED ON THE LEAD. NO ELECTRICAL ANOMALIES WERE NOTED AND THE PATIENT WAS ASYMPTOMATIC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191526 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 86 YR Required Intervention