FDA Adverse Event Injury Summary report: N

CURRENT RF DR

MDR report key: 3090109 · Received May 2, 2013

Report

Report Number
2938836-2013-01097
Event Type
Injury
Date Received
May 2, 2013
Date of Event
February 18, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON THE AVAILABLE PARAMETER AND USAGE INFORMATION, A LONGEVITY CALCULATION WAS PERFORMED AND WAS FOUND TO BE BELOW THE EXPECTED LIMITS. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND NO ANOMALY WAS FOUND. THE CAUSE OF THE PREMATURE BATTERY DEPLETION COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SUDDEN BATTERY DEPLETION WAS OBSERVED. DEVICE TO BE EXPLANTED AND REPLACED.

Description of Event or Problem · 1

NEW INFORMATION RECEIVED NOTES THAT THE DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191293 CURRENT RF DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 2207-36 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention