Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE FOUND AT 12.0-14.5CM FROM THE DISTAL TIP. THE SILICONE INSULATION WAS ABRADED AND THE SENSING CONDUCTOR WAS VISIBLE. OTHER INTERNAL INSULATION ABRASIONS WERE FO UND AT 15.1-16.3CM AND 16.0-17.1CM FROM THE DISTAL TIP. EXTERNAL INSULATION ABRASION WAS FOUND AT 7.3-8.5CM AND 33.0-33.1CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE. THE ETFE COATING WAS INTACT AT ALL ABRASION LOCATIONS.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT DEVICE UPGRADE, EXTERNALIZED CONDUCTORS WERE OBSERVED VIA FLUOROSCOPY. NO ELECTRICAL ANOMALIES WERE NOTED. THE LEAD WAS EXPLANTED AND REPLACED.