FDA Adverse Event Malfunction Summary report: N

UNIFY ASSURA CRT-D

MDR report key: 3090068 · Received May 2, 2013

Report

Report Number
2938836-2013-01239
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
March 4, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED FOR FOLLOW UP. REVIEW OF THE STORED EGMS REVEALED POST-PACED T-WAVE OVERSENSING. DEVICE WAS REPROGRAMMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191004 UNIFY ASSURA CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3257-40 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR