Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

INTERNAL INSULATION ABRASION WAS NOTED AT 9.2-10.3CM FROM THE DISTAL TIP ELECTRODE. THE ETFE COATING WAS INTACT AT THIS LOCATION.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Description of Event or Problem · 1

NEW INFORMATION STATED THAT THE LEAD WAS EXPLANTED DURING THE DEVICE UPGRADE.

Description of Event or Problem · 1

IT WAS REPORTED THAT EXTERNALIZED CONDUCTORS WERE OBSERVED VIA CINE. NO ELECTRICAL ANOMALIES WERE NOTED. LEAD REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192541 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1581/65 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention