FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3090049
·
Received April 19, 2013
Report
- Report Number
- 3008642652-2013-01075
- Event Type
- Malfunction
- Date Received
- April 19, 2013
- Date of Event
- April 11, 2013
- Report Date
- April 16, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE CHARGER/MODEM WOULD NOT POWER ON. UPON EVAL, THE POWER SUPPLY UNIT WAS DEFECTIVE. THE ROOT CAUSE OF THE DEFECTIVE POWER SUPPLY UNIT CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER UNIT. THE LAST PT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, CHARGER/MODEM SN (B)(4) WAS FOUND TO HAVE A DEFECTIVE POWER SUPPLY UNIT. THE LAST PT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 169509 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |