FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3090049 · Received April 19, 2013

Report

Report Number
3008642652-2013-01075
Event Type
Malfunction
Date Received
April 19, 2013
Date of Event
April 11, 2013
Report Date
April 16, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE CHARGER/MODEM WOULD NOT POWER ON. UPON EVAL, THE POWER SUPPLY UNIT WAS DEFECTIVE. THE ROOT CAUSE OF THE DEFECTIVE POWER SUPPLY UNIT CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER UNIT. THE LAST PT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, CHARGER/MODEM SN (B)(4) WAS FOUND TO HAVE A DEFECTIVE POWER SUPPLY UNIT. THE LAST PT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
169509 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA