UNIFY CRT-D, DF4 CONNECTOR

MDR report key: 3090015 · Received May 2, 2013

Report

Report Number
2938836-2013-01220
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
February 27, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. BACKUP VVI RESET MODE WAS VERIFIED IN THE LABORATORY. THE CAUSE WAS DUE TO AN ANOMALOUS COMPONENT WITHIN THE CIRCUITRY.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
191444 UNIFY CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3231-40Q NA