FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 3090004
·
Received April 19, 2013
Report
- Report Number
- 3008642652-2013-01077
- Event Type
- Malfunction
- Date Received
- April 19, 2013
- Date of Event
- April 11, 2013
- Report Date
- April 16, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVAL, THERE WAS A DAMAGED PIN IN THE ELECTRODE BELT TRUNK CABLE CONNECTOR. THE CAUSE FOR THE DAMAGED PIN CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY DUE TO THE CONNECTOR BEING FORCED INTO THE MONITOR WHILE THE PINS WERE MISALIGNED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BELT SN (B)(4) WAS FOUND TO HAVE A DAMAGED PIN IN THE ELECTRODE BELT TRUNK CABLE CONNECTOR. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 169369 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |