ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00330
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- April 4, 2013
- Report Date
- April 4, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT RT212 ADULT INSPIRATORY HEATED BREATHING CIRCUIT IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.
(B)(4). THE RT212 ADULT INSPIRATORY HEATED BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT RT212 ADULT BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A HOLE ABOUT 2.5CM FROM THE CONNECTOR OF THE DRYLINE TUBE. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120316. CONCLUSION: IT WAS IDENTIFIED THAT DAMAGE TO THE RT212 DRYLINE WAS CAUSED BY A FAULTY CORRUGATOR BLOCK THAT WAS USED DURING THE DRYLINE MANUFACTURING PROCESS. THE FAULTY CORRUGATOR BLOCK WAS REPLACED LAST YEAR; HOWEVER, THE SUBJECT DRYLINE TUBE WAS MANUFACTURED PRIOR TO THE REPLACEMENT OF THAT FAULTY BLOCK. DURING THE PRODUCTION OF THE DRYLINE TUBES, A PRESSURE TEST IS PERFORMED (B)(4) AND THOSE THAT FAIL ARE SET ASIDE FOR FURTHER INSPECTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT212 ADULT INSPIRATORY-HEATED BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THERE WAS A HOLE IN ONE OF THE LIMBS OF AN RT212 ADULT INSPIRATORY HEATED BREATHING CIRCUIT. THIS WAS OBSERVED BEFORE USE ON A PATIENT.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THERE WAS A HOLE IN ONE OF THE LIMBS OF AN RT212 ADULT INSPIRATORY HEATED BREATHING CIRCUIT. THIS WAS OBSERVED BEFORE USE ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 184080 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | RT212 | 120316 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |