RESPIRATORY HUMIDIFIER
Report
- Report Number
- 9611451-2013-00329
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- March 31, 2013
- Report Date
- April 4, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K033710
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT MR850AFU RESPIRATORY HUMIDIFIER WILL BE RETURNED TO FPH SERVICE CENTRE IN (B)(6) FOR INSPECTION. WE ARE ALSO IN THE PROCESS OF OBTAINING FURTHER INFORMATION FROM THE HEALTHCARE FACILITY IN ORDER TO ASSIST US WITH THE ANALYSIS AND IDENTIFICATION OF A POSSIBLE ROOT CAUSE OF THE EVENT AS REPORTED. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE INSPECTED THE COMPLAINT DEVICE AND COMPLETED OUR ANALYSIS.
(B)(4). INVESTIGATION DONE ON THE SUBJECT MR850AFU RESPIRATORY HUMIDIFIER. METHOD: THE COMPLAINT MR850AFU WAS RETURNED TO FPH SERVICE CENTRE IN (B)(4) FOR INSPECTION. THE DEVICE WAS VISUALLY INSPECTED AND PERFORMANCE TESTED AS PER MR850 PRODUCT TECHNICAL MANUAL BY A TRAINED FPH SERVICE ENGINEER. OUR ANALYSIS IS ACCORDINGLY BASED ON THE INSPECTION RESULTS AND THE LIMITED INFORMATION PROVIDED BY THE HEALTHCARE FACILITY. RESULTS: THE RETURNED MR850 HUMIDIFIER WAS INSPECTED EXTERNALLY AND NO ISSUES WERE NOTED. IT ALSO PASSED THE PERFORMANCE CHECKS. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120905. CONCLUSION: NO FAULT WAS FOUND WITH THE SUBJECT MR850AFU HUMIDIFIER AS IT PASSED THE VISUAL INSPECTION AND PERFORMANCE CHECK. IT WAS RETURNED TO THE HEALTHCARE FACILITY AFTER THE INSPECTION. INVESTIGATION DONE ON THE BREATHING CIRCUIT KITS USED DURING THE REPORTED INCIDENT. METHODS: TWO BREATHING CIRCUIT KITS, WHICH WERE USED DURING THE REPORTED INCIDENT, WERE RETURNED TO FPH IN (B)(4) FOR INSPECTION. THE FIRST KIT CONTAINED AN RT330 INFANT CONTINUOUS FLOW BREATHING CIRCUIT, AN MR290 HUMIDIFICATION CHAMBER WITH LOT NUMBER 090507, A PRESSURE MANIFOLD, AND A GAS SUPPLY LINE. THE SECOND KIT CONTAINED AN RT330 INFANT CONTINUOUS FLOW BREATHING CIRCUIT, AN MR290 HUMIDIFICATION CHAMBER WITH LOT NUMBER 120816, AN OPT314 OPTIFLOW JUNIOR NASAL CANNULA, A PRESSURE MANIFOLD, AND A GAS SUPPLY LINE. THE RETURNED DEVICES WERE VISUALLY INSPECTED. THE HEATER WIRE RESISTANCE OF THE RT330 INFANT BREATHING CIRCUITS WERE ALSO TESTED USING A MULTIMETER. BOTH BREATHING CIRCUIT KITS WERE SUBSEQUENTLY PERFORMANCE TESTED BY (B)(4). RESULTS: NO PHYSICAL DAMAGE WAS OBSERVED TO THE RETURNED BREATHING CIRCUIT KITS. THE HEATER WIRE RESISTANCE OF THE RT330 INFANT BREATHING CIRCUIT WAS WITHIN SPECIFICATION. NO FAULT WAS FOUND WITH THE BREATHING CIRCUIT KITS DURING PERFORMANCE TEST. NO AUDIBLE OR VISUAL ALARMS WERE OBSERVED ON THE MR850 RESPIRATORY HUMIDIFIERS. CONCLUSION: NO FAULT WAS FOUND WITH THE RETURNED BREATHING CIRCUIT KITS AS IT PASSED THE VISUAL INSPECTION, PERFORMANCE CHECK, AND ELECTRICAL RESISTANCE TEST. BASED ON THE LIMITED INFORMATION WE RECEIVED FROM THE HEALTHCARE FACILITY, WE WERE UNABLE TO DETERMINE DEFINITIVELY THE ROOT CAUSE OF THE REPORTED FAULT OR HOW THE (B)(6) SUSTAINED FIRST DEGREE BURN ON THE CHECK. THE HEALTHCARE FACILITY LATER STATED THAT IT WAS NOT DUE TO THE RT330 OPTIFLOW JUNIOR TUBING. FURTHER INFORMATION RECEIVED REVEALED THAT THE (B)(6) CHEEK HAS SINCE HEALED.
A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT A (B)(6) ALLEGEDLY SUSTAINED A FIRST DEGREE BURN ON THE CHEEK WHILE USING AN MR850AFU RESPIRATORY HUMIDIFIER, AN RT330 OPTIFLOW JUNIOR TUBING KIT, AND A NEOBLENDER. THE (B)(6) WAS ALSO IN AN INCUBATOR WHEN THE INCIDENT OCCURRED. IT WAS FURTHER REPORTED THAT A BIAFINE CREAM WAS APPLIED TO THE BURN AND THAT THE INJURY HAS SINCE HEALED.
A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT A (B)(6) ALLEGEDLY SUSTAINED A FIRST DEGREE BURN ON THE CHEEK WHILE USING AN MR850AFU RESPIRATORY HUMIDIFIER, AN RT330 OPTIFLOW JUNIOR TUBING KIT, AND A NEOBLENDER. THE (B)(6) WAS ALSO IN AN INCUBATOR WHEN THE INCIDENT OCCURRED. IT WAS FURTHER REPORTED THAT A BIAFINE CREAM WAS APPLIED TO THE BURN AND THAT THE INJURY HAS SINCE HEALED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 185441 | RESPIRATORY HUMIDIFIER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR850AFU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | INCUBATOR| NEOBLENDER| MR850 RESPIRATORY HUMIDIFIER| INCUBATOR| RT330 OPTIFLOW JUNIOR TUBING KIT (LOT # 121016)| MR850 RESPIRATORY HUMIDIFIER| RT330 OPTIFLOW JUNIOR TUBING KIT (LOT # 121016)| NEOBLENDER |