FDA Adverse Event Malfunction Summary report: N

COULTER® LH 780 HEMATOLOGY ANALYZER

MDR report key: 3083754 · Received April 29, 2013

Report

Report Number
1061932-2013-00735
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
March 31, 2013
Report Date
March 31, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OBSERVED FLUID LEAKED AT THE TUBING AT PINCH VALVE VL46A. THE FSE REPLACED THE AFFECTED TUBING TO RESOLVE THE FLUID LEAK ISSUE AND REPLACED THE SMALL DIAMETER GREEN TUBING TO THE SHEATH RESTRICTOR COIL AS A PREVENTIVE MEASURE. THE FSE PRIMED THE SYSTEM, VERIFIED SYSTEM STARTUP, LATRON AND CONTROLS. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED LESS THAN TWO MILLILITERS OF FLUID LEAKED WITHIN THE COULTER LH 780 HEMATOLOGY ANALYZER AT THE UPPER RESTRICTOR. THE OPERATOR WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. NO ERRONEOUS PATIENT RESULTS WERE GENERATED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. THE CUSTOMER CLEANED UP THE FLUID LEAK FOLLOWING LABORATORY PROCEDURES; THE LABORATORY HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.