FDA Adverse Event Malfunction Summary report: N

5MM, 33CM PEEK MONOPOLAR HANDLE 33CM

MDR report key: 3083674 · Received April 29, 2013

Report

Report Number
0002936485-2013-00188
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 11, 2013
Report Date
April 11, 2013
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED FAILURE MODE WAS CONFIRMED. DENTS AND SCRATCHES WERE SEEN ON THE INSULATION. ALSO THE INSULATION IS MELTED NEAR THE DISTAL TIP. THE INSULATION WAS TESTED FOR CRACKS/DAMAGES AND FAILED THE INSULATION SCAN TEST. THE PROBABLE ROOT CAUSES ARE USER MISUSE, IMPROPER STERILIZATION METHODS, AND/OR NORMAL WEAR. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION OF THE DEVICE WAS COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION OF THE DEVICE WAS COMPROMISED.