FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3083659 · Received April 29, 2013

Report

Report Number
3004209178-2013-92880
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH BLANK DISPLAY FOLLOWED BY AND UNEXPECTED CONSTANT AUDIO TONE DUE TO FAULTY CAPACITOR ON THE INTERFACE BOARD. DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED. UNIT HAD SCRATCHED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT IS REPORTED THAT THE INSULIN PUMP WAS DROPPED, NOW HAS BLANK DISPLAY AND A LONG CONSTANT TONE. ANOMALY PERSISTS AFTER BATTERY CHANGE. DRIVE SUPPORT CAP IS PROTRUDING. CUSTOMER DID NOT PRESS CAP WHILE CONNECTED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185273 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWS

Patients

Seq Age Sex Outcome Treatment
1