FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3083657 · Received April 29, 2013

Report

Report Number
3004209178-2013-92879
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH ALL BUTTONS RESPONDING PROPERLY. HOWEVER, MOISTURE TRACES WERE FOUND AT KEYPAD TRACES. NO BUTTON ERROR ALARMS WERE NOTED. THE INSULIN PUMP PASSED REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. THE DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED. THE INSULIN PUMP RECEIVED WITH MISSING END CAP STICKER. THE INSULIN PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT IS REPORTED THAT CUSTOMER STATED THAT THE INSULIN PUMP WAS NOT WORKING PROPERLY. BLOOD GLUCOSE READING ARE HIGHER THAN NORMAL. CUSTOMER'S BLOOD GLUCOSE READING IS 19 MMOL/L. CUSTOMER HAS CORRECTED THE BLOOD GLUCOSE LEVEL WITH INJECTION. NO SYMPTOMS OF HIGH BLOOD GLUCOSE. TIME AND DATE ON THE INSULIN PUMP IS CORRECT. PROGRAMMING CORRECT. CUSTOMER MENTIONED THE NOTIFICATION LETTER AND THE DRIVE SUPPORT CAP IS PROTRUDED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185185 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754CAS

Patients

Seq Age Sex Outcome Treatment
1