FDA Adverse Event Malfunction Summary report: N

COULTER® LH500 HEMATOLOGY ANALYZER

MDR report key: 3083578 · Received April 29, 2013

Report

Report Number
1061932-2013-00682
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 3, 2013
Report Date
April 3, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FSE REPLACED THE BLOOD SAMPLING VALVE (BSV) ACTUATOR TO CORRECT THE H/H MISMATCH. THE FSE ALSO FOUND AN AIR LEAK FROM WORN PRIMARY AND SECONDARY ASPIRATION PUMPS, AND REPLACED BOTH PUMPS AS PREVENTIVE MAINTENANCE. PER THE FSE, THE LYSE S III REAGENT WAS ALSO REPLACED TO RESOLVE THE DROP ON HGB RECOVERY. THE FSE SUSPECTED THE REAGENT WAS RECEIVED FROZEN AND MAY HAVE BEEN USED AS SUCH. FAILURE MODE OF THE H/H MISMATCH WAS RELATED TO THE BSV ACTUATOR. THE DROP IN THE HGB RESULTS WAS ATTRIBUTED TO A FROZEN LYSE S III REAGENT, WHICH WOULD MAKE THIS EVENT A USE ERROR. INSTRUCTIONS FOR HANDLING FROZEN REAGENTS ARE DETAILED IN THE PRODUCT IFU. PER IFU INSTRUCTIONS FOR LYSE S III REAGENT PN 8546983; IF PRODUCT HAS BEEN PARTIALLY OR COMPLETELY FROZEN, ALLOW PRODUCT TO WARM TO ROOM TEMPERATURE. MIX PRODUCT BY GENTLE INVERSION PRIOR TO PLACEMENT ON THE INSTRUMENT. INSTALL AND PRIME THE REAGENT AS DIRECTED IN THE INSTRUMENT PRODUCT MANUALS AND/OR ONLINE HELP. VERIFY BACKGROUND COUNTS ARE ACCEPTABLE BEFORE ANALYZING PATIENT SAMPLES. REAGENT INFORMATION: REAGENT: LYSE S III DIFF,1L; PART NUMBER: 8546983; LOT NUMBER: 101515F.

Description of Event or Problem · 1

A CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THEY EXPERIENCED TWO OCCASIONS ON THEIR COULTER LH500 HEMATOLOGY ANALYZER WHERE HEMOGLOBIN (HGB) DROPPED ON CONTROLS (QC) FROM BETWEEN 0.5 AND 1.0 GRAMS. THE CUSTOMER ALSO REPORTED OCCASIONAL HEMATOCRIT/HEMOGLOBIN (H/H) MISMATCH ON SAMPLE RESULTS. NO ERRONEOUS RESULTS WERE GENERATED. THE CUSTOMER STOPPED USING THE INSTRUMENT SINCE THE CONTROLS DID NOT PASS VERIFICATION AND BEC FIELD SERVICE (FSE) WAS DISPATCHED TO THE CUSTOMER SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
184899 COULTER® LH500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 500 NA

Patients

Seq Age Sex Outcome Treatment
1