UV FLASH TRANSFER SETS

MDR report key: 3083428 · Received April 29, 2013

Report

Report Number
1416980-2013-10719
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 5, 2013
Report Date
April 5, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION FOUND A BENT SPIKE AND TUBING THAT WAS CUT OFF NEXT TO THE SLEEVE THAT WAS CUT BY THE CUSTOMER TO SEND THE DEVICE IN. LEAK TESTING WAS PERFORMED WITH NO ISSUES NOTED FOR SLEEVE TO SPIKE ASSEMBLY; CLEAR PASSAGE TESTING AND CLAMP FUNCTION TESTING WERE PERFORMED WITH NO ISSUES NOTED. THE OUTSIDE DIAMETER OF THE UV SPIKE WAS FOUND TO BE WITHIN SPECIFICATIONS. THE REPORTED PROBLEM WAS NOT CONFIRMED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT IT WAS DIFFICULT TO DISCONNECT THE SPIKE OF A TRANSFER SET FROM A PORT OF A TWIN BAG WHEN THE PATIENT USES CLEAN FLASH. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183692 UV FLASH TRANSFER SETS SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1