FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3083272 · Received April 29, 2013

Report

Report Number
1416980-2013-10703
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 4, 2013
Report Date
April 4, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR LOT NUMBER H12H30030 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: DURING VISUAL INSPECTION OF THE RETURNED SAMPLE A CUT IN THE SHEETING WAS OBSERVED. A LEAK AT THE SITE OF THE CUT IN THE SHEETING WAS FOUND DURING LEAK TESTING. CLEAR PASSAGE AND CLAMP FUNCTION TESTS WERE PERFORMED WITH NO ISSUES NOTED. THE CAUSE OF THE LEAK COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A HOME CHOICE CASSETTE, A CUT IN THE CASSETTE SHEETING WAS NOTICED. LEAK TESTING WAS PERFORMED, AND A LEAK WAS NOTED AT THE CUT IN THE SHEETING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185896 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME H12H30030

Patients

Seq Age Sex Outcome Treatment
1 57 YR HOME CHOICE