FDA Adverse Event Summary report: N

RELIANCE 444 WASHER

MDR report key: 3083155 · Received April 29, 2013

Report

Report Number
9680353-2013-00047
Date Received
April 29, 2013
Date of Event
April 5, 2013
Report Date
April 29, 2013
Manufacturer
STERIS CANADA CORPORATION
Product Code
MEC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN ARRIVED AT THE FACILITY AND FOUND THAT THE LEFT HAND RUNNER THAT THE BASKETS SLIDE ON WHEN LOADED INTO THE MACHINE WAS LOOSE AND MISSING ITS SET SCREWS HOLDING THE GUIDE IN PLACE. WHEN A BRACKET WAS RETURNED FROM THE UNLOAD SIDE, IT PUSHED THE RUNNER UP AGAINST THE LOAD DOOR. THE RUNNER THEN CAUGHT A BRACKET LOCATED ON THE LOAD DOOR, PREVENTING THE DOOR FROM OPENING PROPERLY. THE TECHNICIAN REINSTALLED THE SET SCREWS HOLDING THE GUIDE IN PLACE RAN A TEST CYCLE AND CONFIRMED THE UNIT OPERATIONAL. THE TECHNICIAN INFORMED THE OPERATOR NOT TO ATTEMPT TO MANUALLY OPEN THE LOAD DOOR IN THE EVENT IT DOES NOT OPEN TO SPECIFICATION.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT THE LOAD DOOR ON THE RELIANCE WASHER WAS BECOMING STUCK AND WOULD NOT OPEN. THE FACILITY REPORTED AN OPERATOR WAS ATTEMPTING TO OPEN THE LOAD DOOR, WHEN IT BECAME STUCK AT 2 OPEN. THE OPERATOR THEN TRIED TO PUSH THE DOOR UP MANUALLY, AND REPORTED THE DOOR OPENED VERY RAPIDLY, CATCHING HER FINGER BETWEEN THE DOOR AND THE UPPER ACCESS PANEL. THE OPERATOR RECEIVED A SMALL CUT ON HER MIDDLE FINGER, PLACED A BAND-AID ON THE AREA, AND RETURNED TO WORK. THE EMPLOYEE DID NOT SEEK MEDICAL ATTENTION. THERE ARE NO PROCEDURAL DELAYS/CANCELLATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183983 RELIANCE 444 WASHER WASHER MEC STERIS CANADA CORPORATION

Patients

Seq Age Sex Outcome Treatment
1