FDA Adverse Event Injury Summary report: N

SETROX S 60

MDR report key: 3083100 · Received April 29, 2013

Report

Report Number
1028232-2013-01204
Event Type
Injury
Date Received
April 29, 2013
Date of Event
September 6, 2012
Report Date
April 16, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE WERE RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.

Description of Event or Problem · 1

THIS PATIENT WAS NOTED TO HAVE VEGETATION ON THE RV LEAD ON (B)(6) 2012. THIS SYSTEM WAS EXPLANTED DUE TO INFECTION ON (B)(6) 2012. THIS LEAD WAS RETAINED BY THE HOSPITAL. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183717 SETROX S 60 PACER LEAD NVN BIOTRONIK SE & CO. KG 350975

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization