FDA Adverse Event Malfunction Summary report: N

SETROX S 53

MDR report key: 3083070 · Received April 29, 2013

Report

Report Number
1028232-2013-01153
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 9, 2013
Report Date
April 16, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. DURING THIS ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

THIS LEAD HAD TOTAL LOSS OF CAPTURE. THE PHYSICIAN ATTEMPTED TO REPOSITION THIS LEAD BUT IT WOULD NOT REMAIN IN POSITION. THEREFORE, IT WAS EXPLANTED AND REPLACED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185135 SETROX S 53 PACER LEAD NVN BIOTRONIK SE & CO. KG 350974

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization