KAINOX SL 65/16
Report
- Report Number
- 1028232-2013-01147
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- April 17, 2013
- Report Date
- April 17, 2013
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVY
- PMA / PMN Number
- P980023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.
THE RV SHOCK IMPEDANCE HAS BEEN GREATER THEN 150 OHMS SINCE (B)(6) 2013. EXPLANT WAS RECOMMENDED. THE FUTURE DISPOSTION OF THIS LEAD HAS NOT BEEN DECIDED. ON (B)(6) 2013 - AS OF TODAY, ALL AVAILABLE INFORMATION SUGGESTS THIS LEAD REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 185269 | KAINOX SL 65/16 | ICD LEAD | NVY | BIOTRONIK SE & CO. KG | 124236 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Other |