FDA Adverse Event Malfunction Summary report: N

IVT DISPOSABLE

MDR report key: 3083028 · Received April 29, 2013

Report

Report Number
1416980-2013-10679
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 1, 2013
Report Date
April 8, 2013
Manufacturer
BAXTER HEALTHCARE - MALTA
Product Code
KPE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION REVEALED THAT THE BAG HAD A PARTICLE INSIDE WHICH SEEMS TO BE A STRING OF PLASTIC. THE REPORTED CONDITION WAS CONFIRMED. THE ROOT CAUSE WAS NOT DETERMINED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA. SHOULD ADDITIONAL RELEVANT INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER UNITED KINGDOM THAT A 1000ML EVA BAG FOR AUTOMIX HAD FINE THREAD-LIKE FIBERS INSIDE THE BAGS. THERE IS NO REPORT OF PATIENT INVOLVEMENT, INJURY/ADVERSE EVENTS, OR MEDICAL INTERVENTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183495 IVT DISPOSABLE CONTAINER, I.V. KPE BAXTER HEALTHCARE - MALTA 11C31V838

Patients

Seq Age Sex Outcome Treatment
1