FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT ?

MDR report key: 3083014 · Received April 29, 2013

Report

Report Number
2134265-2013-02715
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 1, 2013
Report Date
April 1, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: A VISUAL AND MICROSCOPIC EXAMINATION FOUND THAT THE STENT WAS DAMAGED AT THE PROXIMAL END. STRUTS ON THE MOST PROXIMAL ROW WERE LIFTED UPWARDS. NO FURTHER ISSUES WERE NOTED WITH THE CRIMPED STENT, TIP AND BALLOON OF THE DEVICE. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. TRACES OF BLOOD WERE VISIBLE IN THE GUIDEWIRE LUMEN INDICATING THE DEVICE WAS USED IN VIVO. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A CORONARY STENTING TREATMENT PROCEDURE, STENT DAMAGE WAS NOTED. DURING PREPARATION THEY FOUND THAT PART OF A 2.50X16MM PROMUS ELEMENT STENT WAS LIFTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATION WERE REPORTED AND THE PATIENT STATUS WAS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A CORONARY STENTING TREATMENT PROCEDURE, STENT DAMAGE WAS NOTED. DURING PREPARATION THEY FOUND THAT PART OF A 2.50X16MM PROMUS ELEMENT STENT WAS LIFTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATION WERE REPORTED AND THE PATIENT STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185229 PROMUS ELEMENT ? STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911316250 15472599

Patients

Seq Age Sex Outcome Treatment
1