ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-05526
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Report Date
- April 1, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED A DIM AND DISCOLORED DISPLAY SCREEN. A TEST SCREEN WAS INSTALLED ON THE PUMP AND HAD NO ISSUES. UNRELATED TO THE DISPLAY ISSUE, THE PUMP'S HISTORY SHOWED TIME AND DATE RESETS TO FACTORY DEFAULT. THE BATTERY WAS REMOVED FROM THE PUMP DURING TESTING FOR SIX HOURS, AND THE TIME AND DATE RESET TO FACTORY DEFAULT. DURING INVESTIGATION, THE INTERNAL BATTERY WAS FOUND TO BE LEAKING. ADDITIONALLY, EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM AND DISCOLORED DISPLAY SCREEN. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2013. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 183488 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 8 YR |