FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3082993 · Received April 29, 2013

Report

Report Number
1416980-2013-10674
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 7, 2013
Report Date
April 7, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER THE SAMPLE WAS NOT AVAILABLE AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION OR IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN TUBING) ALARM OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 3 OF 4 WITH THE HOME PATIENT (HP) CONNECTED. ONE OF THE SUPPLY BAGS HAD FALLEN AND HAD DISCONNECTED. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) HELPED THE HP TO CLEAR THE ALARM BY TURNING THE HC OFF AND ON AND THE HP WOULD FINISH WITH MANUAL BAGS. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
184142 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 64 YR HOME CHOICE