FDA Adverse Event
Malfunction
Summary report: N
LINOX SD 65/18
MDR report key: 3082861
·
Received April 29, 2013
Report
- Report Number
- 1028232-2013-01119
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- April 10, 2013
- Report Date
- April 15, 2013
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVY
- PMA / PMN Number
- P980023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.
Description of Event or Problem · 1
THIS LEAD WAS CAPPED AND REPLACED DUE TO LOW IMPEDANCES AND ELEVATED THRESHOLDS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 183751 | LINOX SD 65/18 | ICD LEAD | NVY | BIOTRONIK SE & CO. KG | 350054 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Hospitalization |