FDA Adverse Event Injury Summary report: N

LOCKING SCREW, FULLY THREADED S2 Ø5X40 MM

MDR report key: 3082659 · Received April 29, 2013

Report

Report Number
0009610622-2013-00222
Event Type
Injury
Date Received
April 29, 2013
Date of Event
March 10, 2013
Report Date
March 15, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-KIEL
Product Code
HSB
PMA / PMN Number
K032579
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED ISSUE WAS CONFIRMED. DURING INVESTIGATION NO MATERIAL, DESIGN OR MANUFACTURING RELATED ISSUES WERE FOUND, NEITHER FOR THE GAMMA3 LONG NAIL NOR FOR THE LOCKING SCREW. THE EVALUATION REVEALED THAT THE LONG NAIL AS WELL AS THE 40MM LOCKING SCREW BROKE IN A FATIGUE FRACTURE DUE TO OVERLOAD. NO DISCREPANCIES WERE DETECTED DURING REVIEW OF RISK ANALYSIS. THE LOCKING SCREW WITH CAT. # 17965050S IS REGARDED AS CONCOMITANT PRODUCT, AND THE EVENT OF SCREW BREAKAGE HAS TO BE EXAMINED IN CORRELATION WITH THE BREAKAGE OF THE LONG NAIL KIT R2.0, STST, LEFT GAMMA3 Ø11X340MM X 125° WITH CAT. # 43250340S. THUS, THE EVENT IS REGARDED NOT DEVICE RELATED. DUE TO MISSING INFORMATION SUCH AS PERIOD OF IMPLANTATION, ANY PATIENT DATA AND - ANAMNESIS, X-RAYS PRE-, INTER- AND POSTOPERATIVE AS WELL AS SURGICAL RECORDS, SURGEONS¿ INSTRUCTIONS AND PATIENTS¿ ACTIVITY LEVEL A MORE PRECISE STATEMENT REGARDING THE ROOT CAUSE OF THE EVENT IS NOT POSSIBLE.

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THE SURGEON REPORTED TO THE SALES REP, THAT "THE GAMMA 3 NAIL FRACTURED AND WAS EXPLANTED."

Description of Event or Problem · 1

THE SURGEON REPORTED TO THE SALES REP, THAT "THE GAMMA 3 NAIL FRACTURED AND WAS EXPLANTED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183742 LOCKING SCREW, FULLY THREADED S2 Ø5X40 MM IMPLANT HSB STRYKER OSTEOSYNTHESIS-KIEL K121707

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention