FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC CLIP APPLIER

MDR report key: 3082424 · Received April 29, 2013

Report

Report Number
3005075853-2013-02015
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
March 29, 2013
Report Date
April 3, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION IS UNAVAILABLE; DEVICE WAS NOT RETURNED FOR EVALUATION. ADDITIONAL INFORMATION: MULTIPLE REQUESTS FOR RETURN OF DEVICE HAVE BEEN MADE BUT NO DEVICE HAS BEEN RETURNED.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY PROCEDURE, THE CLIP WAS NOT DELIVERED BUT THE APPLIER CUT THE CYSTIC ARTERY. THERE WAS BLEEDING BUT NO TRANSFUSION. THE SURGERY WAS DIFFICULT BECAUSE OF THE CIRRHOSIS OF THE PATIENT AND THIS ISSUE MAKES THE SURGERY MORE DIFFICULT. UNKNOWN HOW CASE WAS COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185221 LIGACLIP** ENDOSCOPIC CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA K4C42N

Patients

Seq Age Sex Outcome Treatment
1