FDA Adverse Event Malfunction Summary report: N

KNIFE

MDR report key: 3082284 · Received April 24, 2013

Report

Report Number
2523835-2013-00053
Event Type
Malfunction
Date Received
April 24, 2013
Date of Event
December 17, 2012
Report Date
March 25, 2013
Manufacturer
ALCON PRECISION DEVICE - SINKING SPRING
Product Code
HNN
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO SAMPLES WERE RETURNED FOR EVALUATION; THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TESTY VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO MANUFACTURER'S ACCEPTANCE CRITERIA. BECAUSE A SAMPLE WAS NOT RETURNED AND NO EVIDENCE OF NONCONFORMITY COULD BE FOUND IN THE LOT RECORD REVIEW, THE ROOT CAUSE FOR THE DULL KNIFE EXPERIENCED BY THE CUSTOMER CANNOT BE DETERMINED. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. ANY DEFECTS, SUCH AS DAMAGED TIPS AND CUTTING EDGES, ARE REMOVED FROM THE LOT AND SCRAPPED. PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. (B)(4).

Description of Event or Problem · 1

A MEDWATCH FORM WAS RECEIVED INDICATING THAT A CUSTOMER REPORTED A KNIFE BLADE WAS ATTEMPTED TO BE USED DURING SURGERY, BUT THE BLADE WAS DULL. THERE WAS NO PATIENT HARM IDENTIFIED ON THE RECEIVED REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
176039 KNIFE MANUAL OPHTHALMIC SURGICAL INSTRUMENT HNN ALCON PRECISION DEVICE - SINKING SPRING 8065982465 904210M

Patients

Seq Age Sex Outcome Treatment
1 55 YR